Automatic Cold Filter Plugging Point Tester: RCF-01CML / -02CML / -03CML / -04CML / -06CML

This is a dedicated cold filter plugging point tester for laboratories that do not require combined pour and cloud point measurement.

Rigo Automatic Cold Filter Plugging Point Tester: RCF-01CML / -02CML / -03CML / -04CML / -06CML — Pour Point, Cloud Point & CFPP Testers

Manufacturer: Rigo — Japan

Overview

This is a dedicated cold filter plugging point tester for laboratories that do not require combined pour and cloud point measurement. A lower photo-sensor that detects whether a drawn-up sample has failed to return to the test tube is fitted as standard, providing the same reverse-plugging protection found on RIGO's combined instruments. Each test bath can be controlled independently, following either a staged temperature program or a linear temperature gradient. In the staged mode, the two-refrigerator circulation system steps the bath temperature through 48°C, -34°C, -51°C, and -67°C in sequence, while the linear gradient mode can run at a rate expressed in °C per minute or, for field-style slow cooling, in °C per hour.

Reverse Plugging Point Detection

A dedicated lower photo-sensor confirms whether the sample drawn into the pipette has returned to the test tube; if it fails to return within the expected time, the instrument records that condition as part of the CFPP result. This standard feature reduces the risk of a false reading when highly waxy samples resist flowing back through the filter.

Staged or Linear Temperature Control

Operators can choose RIGO's proprietary two-bath circulation step program, which automatically shifts the test bath through 48°C, -34°C, -51°C, and -67°C without removing the sample container, or a linear cooling gradient set in °C per minute. A slower linear gradient expressed in °C per hour is also available, intended primarily for field testing conditions.

Common Specifications

SpecificationRCF-01CML / -02CML / -03CML / -04CML / -06CML
Applicable standardsJIS K2288 / ASTM D6371 / IP 309
Measuring range45 to -60°C
Temperature detectionPlatinum resistance thermometer (Pt 100Ω)
Meniscus detectionPhoto-electric detection
Reverse plugging point detectionPhoto-electric detection
Measuring pitch1°C
Cooling methodDual-refrigerator circulation type (automatic coolant circulation)
Staged temperature controlAutomatic 4-step control: 48°C, -34°C, -51°C, -67°C
Linear gradient control°C per minute or °C per hour
Temperature controlProportional control and heater On/Off

Model Comparison

SpecificationRCF-01CMLRCF-02CMLRCF-03CMLRCF-04CMLRCF-06CML
No. of test positions12346