Dynamic Light Scattering Instrument: Litesizer DLS

Litesizer DLS delivers best-in-class particle sizing with automatic angle selection to prevent errors, while MAPS technology yields the highest-possible peak resolution.

Anton Paar Dynamic Light Scattering Instrument: Litesizer DLS — Particle Size Analyzers

Manufacturer: Anton Paar — Austria

Overview

Litesizer DLS delivers best-in-class particle sizing with automatic angle selection to prevent errors, while MAPS technology yields the highest-possible peak resolution. Continuous transmittance monitoring detects sedimentation and agglomeration during measurement, enhancing measurement reliability. Unique to our dynamic light scattering instrument, cmPALS and the Omega Cuvette improve zeta potential accuracy and repeatability by addressing aging effects and minimizing electrical gradient. Our dynamic light scattering instruments also offer fluorescence and polarization filters, not only for particle size measurements, but also for particle concentration in both MAPS three-angle and single-angle modes.

Three Angles, Real-Time Monitoring for Better Sizing

Litesizer DLS offers the choice between three different detection angles to measure a wide array of samples, while eliminating errors with automatic angle selection. The Multi-Angle Particle Sizing (MAPS) measurement mode delivers exceptional resolution, supporting advanced analytical precision. Unique to our DLS particle size analyzers, continuous transmittance monitoring provides live feedback on sample behavior throughout measurement, giving users instant information on sedimentation or agglomeration. This real-time insight enhances measurement quality and reliability, making our dynamic light scattering instrument the optimal choice for accurate and dependable particle analysis.

Market-Leading Zeta Potential Analysis

Our dynamic light scattering instrument's patented cmPALS technique is specially designed to eliminate reproducibility uncertainties caused by aging effects. Additionally, the Omega Cuvette's innovative design minimizes the gradient in the electrical field, further enhancing reproducibility (±3 %). With an industry-leading particle size range for zeta potential measurement, from 1.3 nm to 100 µm, our dynamic light scattering instrument offers the broadest capability available, ensuring precise and consistent results across a vast array of applications.

Kalliope: The Benchmark Software for Particle Sizing

The Kalliope software requires very little user training. Measurements with the dynamic light scattering instrument can be performed with only three clicks. The one-of-a-kind one-page layout provides an instant overview of input parameters, measurement signals, and results, presenting everything you need at a glance. Quickly access measurement results with predefined standard reports and organized result summaries. For more detailed analysis, use the customizable report templates, Excel export options, and advanced analysis functions. Kalliope is also fully compliant with 21 CFR Part 11, ensuring robust data integrity and regulatory adherence.

All the Measuring Modes You Need

Enhance your versatility with Litesizer DLS 701's particle concentration measurement mode, enabling calibration-free concentration analysis for up to three distinct size populations within a single sample. This mode supports both monodisperse and polydisperse samples across a wide concentration range, whether you're using single-angle DLS or MAPS. You can additionally measure the refractive index of solvents at the exact wavelength and temperature of your measurements (European Patent 3 023 770), ensuring maximum accuracy for particle size and zeta potential results. Litesizer DLS 701 and DLS 501 provide precise refractive index measurements within ±0.5 %. Our dynamic light scattering instrument also offers molecular mass measurement.

Most Capable Filtering Options Available

Litesizer DLS 701 and 501 are the only dynamic light scattering instruments on the market that offer users the choice between fluorescence or polarization filters, which can be applied in all three angle directions. This enables particle concentration analysis as well as MAPS three-angle or selected single-angle analysis with applied filters – permitting types of analysis not possible with comparable instruments (e.g., concentration determination of quantum dots).

10-Year Laser Warranty: Confidence, for a Decade

Each Litesizer DLS is built around a high-performance laser system engineered for long-term stability and reliable operation. We make sure our instruments' core optical performance supports precise, reproducible DLS-based particle size measurements for many years, regardless of application demands. That's why we offer a 10-year warranty on the laser – a commitment that goes far beyond industry standards. We'll win your confidence, for a decade.

Models

Dynamic Light Scattering Instrument: Litesizer DLS is available in 4 models: Litesizer DLS 701, Litesizer DLS 501, Litesizer DLS 301, Litesizer DLS 101.

Litesizer DLS 701

Particle sizing from 0.3 nm to 15 µm.

Litesizer DLS 501

Particle sizing from 0.3 nm to 15 µm.

Litesizer DLS 301

Particle sizing from 0.3 nm to 15 µm.

Litesizer DLS 101

Particle sizing from 0.3 nm to 10 µm.

Specifications

SpecificationLitesizer DLS 701Litesizer DLS 501Litesizer DLS 301Litesizer DLS 101
Particle Size Specifications
Measurement PrincipleDynamic Light Scattering (DLS)
Measuring Range0.3 nm to 15 μm* (particle diameter)0.3 nm to 15 μm* (particle diameter)0.3 nm to 15 μm* (particle diameter)0.3 nm to 10 μm* (particle diameter)
Measurement Angles15°, 90°, 175°, Multi-angle particle sizing (MAPS)15°, 90°, 175°90°175°
Min. Concentration0.1 mg/mL (lysozyme)
lower than 0.00001 % (0.1 ppm, Latex 100 nm)
0.1 mg/mL (lysozyme)
lower than 0.00001 % (0.1 ppm, Latex 100 nm)
1 mg/mL (lysozyme)0.1 mg/mL (lysozyme)
Max. Concentration50 % w/v (sample-dependent)50 % w/v (sample-dependent)40 % w/v (sample-dependent)50 % w/v (sample-dependent)
Min. Sample Volume1.5 µL1.5 µL1.5 µL12 µL
AccuracyBetter than ±2 % on NIST traceable standards
RepeatabilityBetter than ±2 % on NIST traceable standards
Particle Concentration Specifications
Analysis ModelMie Theory---
Measuring Range108 - 1013 particles/mL (sample dependent)---
Size Limit1 µm---
Measurement Angles15°, 90°, 175°, Multi-angle particle sizing (MAPS)---
Min. Sample Volume12 µL---
Accuracy± 10 % (sample dependent)---
Repeatability± 5 % (sample dependent)---
Zeta Potential Specifications
Measuring PrincipleElectrophoretic Light Scattering (ELS) / cmPALSElectrophoretic Light Scattering (ELS) / cmPALSElectrophoretic Light Scattering (ELS) / cmPALS-
Measuring Range>= ±1000 mV>= ±1000 mV>= ±1000 mV-
Mobility Range10⁻¹¹ m²/V.s to 2 x 10⁻⁷ m²/V.s10⁻¹¹ m²/V.s to 2 x 10⁻⁷ m²/V.s10⁻¹¹ m²/V.s to 2 x 10⁻⁷ m²/V.s-
Size Range1.3 nm to 100 µm (diameter)1.3 nm to 100 µm (diameter)1.3 nm to 100 µm (diameter)-
Min. Sample Concentration0.1 mg/mL (lysozyme)0.1 mg/mL (lysozyme)0.1 mg/mL (lysozyme)-
Max. Sample Concentration70 %w/v (sample-dependent)70 %w/v (sample-dependent)70 %w/v (sample-dependent)-
Max. Sample Conductivity200 mS/cm200 mS/cm200 mS/cm-
Min. Sample Volume50 µL (sample-viscosity-dependent)50 µL (sample-viscosity-dependent)50 µL (sample-viscosity-dependent)-
AccuracyBetter than ±10 %Better than ±10 %Better than ±10 %-
Repeatability±3 %±3 %±3 %-
Molecular Mass Specifications
Measuring PrincipleStatic Light Scattering (SLS)Static Light Scattering (SLS)Static Light Scattering (SLS)-
Measuring Range (Mass)300 Da to 20 MDa300 Da to 20 MDa300 Da to 20 MDa-
Measuring Range (Particle Size)Up to 40 nm (diameter)Up to 40 nm (diameter)Up to 40 nm (diameter)-
Measurement Angle90°90°90°-
Min. Sample Concentration0.1 mg/mL (lysozyme)0.1 mg/mL (lysozyme)0.1 mg/mL (lysozyme)-
Accuracy±10 %±10 %±10 %-
Repeatability±5 %±5 %±5 %-
Transmittance Specifications
Measuring Range0 % to 100 %
Min. Sample Volume15 µL
AccuracyBetter than ±1 %
Refractive Index Specifications
Measuring Range1.28 to 1.501.28 to 1.50--
Min. Sample Volume1 mL1 mL--
Accuracy±0.5 %±0.5 %--
General Specifications
Light SourceSemiconductor laser diode / 40 mW, 658 nm
Optical Filters (Fluorescence, Vertical and Horizontal Polarization)forward, side and backscatteringforward, side and backscatteringforward, side and backscattering-
Laser Warm-Up Time6 min
Temperature Control Range0 °C to 120 °C
Ambient Operating Temperature10 °C to 35 °C
HumidityUp to 80 % non-condensing
Dimensions (W, D, H)450 mm, 505 mm, 135 mm
WeightApprox. 18 kg (40 lb)Approx. 18 kg (40 lb)Approx. 18 kg (40 lb)Approx. 16.3 kg (36 lb)

Interested in this model? Let's talk specifics.

Request a formal quotation or a detailed technical discussion — our engineers respond within one business day.

Request a Quote