Manage your lab with efficiency: The Abbemat Advanced series laboratory refractometers achieve precise, consistent measurements in busy and automated labs. Conduct sample-specific measurements according to your SOPs. Implement the instruments into multi-parameter measuring systems, and equip them with an autosampler to avoid errors and boost your lab efficiency. Straightforward integration of these laboratory refractometers into IT networks and data management via Anton Paar Connect ensures data integrity and fast access in paperless labs.

Complete Qualification of your Laboratory Refractometer

Performance qualification and regular control of measuring equipment does not stop at verification of the refractive index. Anton Paar also includes the second important measuring parameter: temperature. Correct temperature readings of your automatic refractometer are confirmed by Anton Paar service engineers with the patented T-Check system:

  • Traceable temperature check and adjustment on-site
  • Most precise prism-surface temperature measurement
  • Prevention of temperature-induced errors
  • Documentation and certificate

Error Prevention via User Guidance

User rights management ensures operators only get access to the functionality they need for their job. Settings can only be changed with corresponding privileges.

Routine measurements are organized by sample/product, with locked settings and pass/fail criteria defined by your SOPs.

Check-intervals can be set to prove system suitability.

Networked Data Management Options

Abbemat laboratory refractometers can export file-based data straight to defined network folders.

The Abbemat Advanced series, like other instruments from Anton Paar, can be connected to our lab execution software AP Connect. It provides simple and secure data workflows in laboratories for single or multiple instruments via a centralized digital hub. This ensures paperless efficiency and data integrity in a central MS-SQL database.

Efficiency via Combination with other Anton Paar Instruments and Laboratory Solutions

Abbemat Advanced laboratory refractometers blend into automated and networked laboratory workflows and improve their efficiency:

  • Auto-sampler
  • Multi-parameter measuring systems
  • Over 200 pre-installed methods for quick concentration measurements
  • Flow cells and funnels for fast sample application
  • Configurable screen to focus only on what is needed
  • Measurement cycle times down to 10 seconds

Scratch-Proof Sapphire Prism with 25-Year Warranty

Designed to last, top-grade optical components like the sapphire prism ensure ruggedness and a long instrument life of your automatic laboratory refractometer.

  • Resists cleaning with spatula or other lab tools
  • Mohs hardness of 9, close to diamond-grade
  • Backed by a 25-year warranty on the prism

Models

Laboratory Refractometer: Abbemat Advanced is available in 3 models: Abbemat Advanced 5001, Abbemat Advanced 5101, Abbemat Advanced 5201.

Abbemat Advanced 5001

Accuracy: ±0.0001 nD / ± 0.05 °Brix.

Abbemat Advanced 5101

Accuracy: ±0.00004 nD / ± 0.03 °Brix.

Abbemat Advanced 5201

Accuracy: ±0.00002 nD / ± 0.015 °Brix.

Specifications

SpecificationAbbemat 5001Abbemat 5101Abbemat 5201
Refractive Index [nD]
Range1.33 nD to 1.55 nD1.30 nD to 1.66 nD1.26 nD to 1.72 nD
Max. Resolution0.0001 nD0.00001 nD0.000001 nD
Accuracy¹⁾±0.0001 nD±0.00004 nD±0.00002 nD
Repeatability±0.0001 nD±0.0000015 nD±0.0000015 nD
Brix
Range0 °Brix to 100 °Brix
Display Resolution0.01 °Brix0.001 °Brix0.001 °Brix
Accuracy±0.05 °Brix±0.03 °Brix±0.015 °Brix
Temperature
Range20 °C to 40 °C10 °C to 70 °C4 °C to 105 °C ²⁾
Probe Accuracy¹⁾±0.05 °C±0.03 °C±0.03 °C
Probe Stability±0.002 °C
Measuring Specifications
Display7" LCD touch screen
Communication Interfaces1 x Ethernet
3 x USB B
CAN-Bus for connection with DMA density meter, MCP polarimeter, SVM viscometer, Xsample 520 sample changer
FeaturesUser management, sample-specific configurations, pass/fail, preconfigured methods, AP Connect ready
Data Storage10,000 measuring results
400 products
General
Min. Sample Volume70 µL
Max. Allowed Pressure in Micro Flow Cell10 bar
Ambient Temperature5 °C to 40 °C
Materials in Contact with the Sample
PrismSynthetic sapphire
Sample WellStainless steel, optionally Ni-alloy
SealFFKM (perfluoroelastomer)
Compounds
Light SourceLED light source, average lifetime > 100,000 hours
Wavelength589.3 nm (by wavelength-adjusted interference filter)
Power Requirements100-240 VAC +10 %/-15 %, 50/60 Hz, min. 10 W, max. 100 W, depending on sample temperature setting and ambient temperature
Dimensions
W x H x D225 mm x 115 mm x 330 mm
Weight3.5 kg
Notes¹⁾ Valid at standard conditions (T=20 °C, ambient temperature 20 °C to 25 °C
²⁾ At max. ambient temperature of 35 °C