High Vacuum Physisorption Analyzer: Autosorb 6100, 6200, and 6300

High-vacuum physisorption and chemisorption analyzers for advanced measurement of BET surface area, active metal area, and pore size distribution — independent analysis stations with dedicated transducers deliver research-grade isotherms with maximum throughput.

Anton Paar High Vacuum Physisorption Analyzer: Autosorb 6100, 6200, and 6300 — Surface Area Analyzers

Manufacturer: Anton Paar — Austria

Overview

Autosorb instruments are high-vacuum physisorption and chemisorption analyzers designed for advanced measurement of BET surface area, active area, and pore size distribution. Independent analysis stations, exclusive TruZone active coolant control, and user-friendly Kaomi software satisfy all of your measurement needs, while easy-to-install accessory kits empower you to adapt to future ones. The choice is clear: Absolutely Autosorb.

Your Gas Sorption Analyzer: Absolutely Accurate

  • Perform highly accurate physisorption measurements from relative pressures below 10⁻⁷ for nitrogen
  • Leverage precise manifold temperature control between 35 °C and 50 °C
  • Rely on TruZone active coolant level control for accurate data – even throughout lengthy analyses
  • Achieve superior high-vacuum performance (38 % better than previous-generation instruments) with precision-machined manifolds and high-quality pneumatic valves

Independent Stations: Absolutely Agile

  • Analyze up to three different samples with three different analysis gases at three different temperatures simultaneously with independent analysis stations and patented temperature control accessories
  • Perform the most challenging physisorption measurements with the 90-hour Dewar
  • Maximize your throughput with six integrated degassing stations
  • Quickly swap to the 1,100 °C furnace for advanced temperature-programmed analysis, pulse titrations, or chemisorption isotherms

Intuitive Software: Absolutely Accessible

  • Enjoy the updated, streamlined Kaomi software, which makes the whole measurement process easier than ever before
  • Optimize measurement definitions to suit your needs with the new standard and advanced modes of DoseWizard
  • Keep downtime to a minimum with PowderProtect to prevent accidental loss of powder in the instrument

Customization Options: Absolutely Adaptable

  • With three instruments to choose from, Autosorb adapts to the exact analysis requirements of your porous solids and catalysts
  • As your research changes, accessory kits covering physisorption, chemisorption, vaporsorption, and more can be added to fit new materials and applications
  • Go beyond standard measurements with enhanced chemical resistance, a built-in vapor source, a thermal conductivity detector, or even a fully integrated mass spectrometer

Compliance and Quality: Absolutely Assured

  • Comply with 20+ ASTM, DIN, and ISO standards for catalyst characterization and the evaluation of porous solids
  • Obtain peace of mind with our three-year warranty so you can avoid unforeseen costs

Models

High Vacuum Physisorption Analyzer: Autosorb is available in 3 models: Autosorb 6100, Autosorb 6200, Autosorb 6300.

Autosorb 6100

Accurate high-vacuum physisorption analysis of surface area and micropores.

Autosorb 6200

Accurate physisorption and chemisorption analysis for advanced catalyst characterization.

Autosorb 6300

Full chemisorption and physisorption capabilities in a single instrument package.

Specifications

SpecificationAutosorb 6100Autosorb 6200Autosorb 6300
Measurement Specifications
PrincipleVacuum volumetricVacuum volumetricVacuum volumetric dynamic flow
Adsorption GasesStandard
N₂, Ar, Kr, CO₂, CO, H₂, CH₄, C₂H₆, C₃H₈
Extended range
N₂, Ar, Kr, CO₂, CO, H₂, CH₄, C₂H₆, C₃H₈, NH₃
Extended range
N₂, Ar, Kr, CO₂, CO, H₂, CH₄, C₂H₆, C₃H₈, NH₃
Adsorption VaporsN/AH₂O, C₆H₆, C₅H₅N, C₃H₆O, C₂H₅OH, CH₃OH, C₃H₇OHH₂O, C₆H₆, C₅H₅N, C₃H₆O, C₂H₅OH, CH₃OH, C₃H₇OH
Physisorption Analysis StationsNumber: 1, 2, or 3
Independent analysis stations: Up to three gases at three analysis temperatures can be used concurrently (one analysis gas and temperature per station)
Independent p0 station (dedicated cell and transducer)
Chemisorption Analysis StationsN/A11
Maximum Pressure1,100 Torr (0.997 p/p0 for N₂ 77 K)
Pressure Resolution2 × 10⁻⁵ Torr (2.6 × 10⁻⁸ p/p0 for N₂ 77 K)1 × 10⁻⁶ Torr (1.3 × 10⁻⁹ p/p0 for N₂ 77 K)1 × 10⁻⁶ Torr (1.3 × 10⁻⁹ p/p0 for N₂ 77 K)
BET Surface Area- Absolute detection limit: 0.1 m² (N₂ 77 K)
- Specific detection limit: 0.01 m²/g (N₂ 77 K)
- Typical reproducibility: 1 % (measured on BAM P115)
- Reproducibility limit: 2 % with 2 m² in the cell
Pore Size- Range: 0.35 nm to 500 nm (diameter)
- Typical reproducibility: 0.5 % (measured on BAM P115)
Active AreaN/A- Absolute detection limit: 0.03 m² (H₂ on platinum 313 K)
- Specific detection limit: 0.003 m²/g (H₂ on platinum 313 K)
- Typical reproducibility: 2 % (measured on 2 % platinum on alumina reference sample)
- Absolute detection limit: 0.03 m² (H₂ on platinum 313 K)
- Specific detection limit: 0.003 m²/g (H₂ on platinum 313 K)
- Typical reproducibility: 2 % (measured on 2 % platinum on alumina reference sample)
TruZoneYes (active coolant level control)
PowderProtectYes (prevents sample elutriation)
Analysis Dewar- Duration: 90+ hours with liquid nitrogen
- Refill during analysis: Yes
- Volume: 3 L
Analysis FurnaceN/A- Maximum temperature: 1,100 °C
- Maximum ramp rate: 50 °C per minute
- Furnace cooling by built-in fan
- Maximum temperature: 1,100 °C
- Maximum ramp rate: 50 °C per minute
- Furnace cooling by built-in fan
Sample Preparation- Six integrated degassing stations with dedicated cold trap
- Temperature control: two independent heating zones, ambient to 450 °C
- Available methods: flow and vacuum, programmable multi-step degassing profiles, test for completion routines, pressure-controlled heating
CryoSync ControlN/AYesYes
Mass SpectrometerN/AN/AYes
Software Specifications
Kaomi for Autosorb- Control up to four instruments from one PC
- Eight languages: Chinese, English, French, German, Japanese, Korean, Portuguese, Spanish
- DoseWizard: 45+ built-in analysis profiles (ASTM, USP, DIN, ISO)
Technical Specifications
Dimensions (W x D x H)79.5 cm × 70.1 cm × 107.9 cm (31.3 in × 27.6 in × 42.5 in) depending on instrument configuration
Weight136.4 kg (300 lbs) depending on instrument configuration
Operating Environment- Temperature: 15 °C to 35 °C (59 °F to 95 °F)
- Humidity: 20 % RH to 80 % RH, non-condensing
- Indoor use only
- Altitude up to 3,000 m
Analysis Manifold ConstructionVacuum-brazed manifold block in stainless steel
Analysis Manifold TemperatureUser-adjustable 35 °C to 50 °C (stability: ±0.05 °C)
Pressure Transducer (1,550 Torr)- Number: 3 to 5 (depending on instrument configuration)
- Accuracy: ±0.1 % of span (typical), ±0.15 % of span (maximum)
- A/D data acquisition: 32 bit
Pressure Transducer (10 Torr)- Number: 1 to 3 (depending on instrument configuration)
- Accuracy: ±0.15 % reading up to 10 Torr
- Resolution: 0.002 % of full scale
Pressure Transducer (1 Torr)- Number: 1 to 3 (depending on instrument configuration)
- Accuracy: ±0.15 % reading up to 1 Torr
- Resolution: 0.002 % of full scale
N/AN/A
Pressure Transducer (0.1 Torr)N/A- Number: 1 to 3 (depending on instrument configuration)
- Accuracy: ±0.15 % reading up to 0.1 Torr
- Resolution: 0.001 % of full scale
- Number: 1 to 3 (depending on instrument configuration)
- Accuracy: ±0.15 % reading up to 0.1 Torr
- Resolution: 0.001 % of full scale
Pirani Vacuum Gauge- Number: 1
- Accuracy:
- 5 × 10⁻⁴ Torr to 1 × 10⁻³ Torr: ±10 % of reading
- 1 × 10⁻³ Torr to 100 Torr: ±5 % of reading
- 100 Torr to atmosphere: ±25 % of reading
Vacuum System- Turbo-molecular drag pump and dry-diaphragm pump
- Manufacturer’s specification: 5 mbar × 10⁻¹⁰ mbar to 2.67 mbar × 10⁻⁵ mbar
- Typical pressure at analysis port: 2.67 mbar × 10⁻⁵ mbar
Thermal Conductivity Detector (TCD)N/AN/A- Two-filament TCD with rhenium/tungsten filaments
- Detection Limit: 0.5 µL (air injected into helium)
Input Ports- 7
(five analysis (FKM), one helium, one degas/backfill)
- 8
(four analysis (FKM), one analysis (PFE), one vapor, one helium, one degas/backfill)
- 16 ports
(11 analysis (FKM), one analysis (PFE), one vapor, one helium, one degas/backfill, one titration)
Gas Supply- Purity: 99.999 %
- Input pressure: 8 psig to 10 psig (0.55 bar to 0.69 bar)
Air- Input pressure: 50 psig to 100 psig (3.45 bar to 6.9 bar)
Electrical- Supply: AC 100 to 240 V AC, 50 / 60 Hz
- Maximum consumption: 1,810 VA (depending on instrument configuration)
PC ConnectionEthernet
RoHS3 CompliantYes
CE / UKCA CertifiedYes