Type-Approved Density Meter: DMA
Anton Paar's type-approved density and concentration meters deliver legally compliant results wherever measurements must stand up to official inspection — designed for regulated industries with outstanding precision, automation, and application expertise.

Manufacturer: Anton Paar — Austria
Overview
Anton Paar’s type-approved density and concentration meters deliver legally compliant results wherever measurements need to stand up to official inspection. Designed for regulated industries, they combine decades of application expertise with outstanding precision, automation, and connectivity. From incoming goods to final quality release, you obtain traceable, auditable data in a single, highly efficient workflow – ready for legal-for-trade use and your digital laboratory.
Type-Approved Accuracy for Legal Metrology
Rely on density and concentration results that are ready for authorities, auditors, and customers. The DMA EZ series is type-approved and designed specifically for legal metrology, so your measurements achieve official calibration and verification. Combined with Anton Paar’s decades of know-how, this secures full confidence in product release decisions and simplifies compliance with national and international standards.
Combine for a Comprehensive Measurement System
Create exactly the measuring system you need today and expand it tomorrow. By combining DMA 4101 EZ, 4501 EZ, and 5001 EZ with various Anton Paar measuring modules, you determine all relevant parameters – such as density, concentration, and more – in a single measurement run. This modularity reduces sample handling, minimizes operator influence, and shortens overall analysis time, while ensuring a consistent, harmonized workflow across different products, methods, and laboratory locations.
Automation and Digital Connectivity
Boost throughput and reduce manual steps with automated sample changers that handle filling, cleaning, and measurement sequences. This improves safety and frees your team to focus on higher-value tasks. Direct integration into the AP Connect lab execution system brings all results into one digital platform, enabling paperless documentation, easy tracking of legal-for-trade measurements, and seamless data transfer to LIMS or ERP systems for end-to-end compliance.
U-View and FillingCheck: Error-Free Measurements
See what’s happening inside the measuring cell – in real time. U-View provides a zoomable live image so operators can visually verify every filling, while FillingCheck automatically detects microbubbles or improper filling conditions. Together they significantly reduce measurement errors and rework, enabling secure, traceable results even with challenging or opaque samples, and giving operators immediate feedback that supports robust, standardized routines.
Models
Type-Approved Density Meter: DMA is available in 5 models: DMA 501 EZ, DMA 1001 EZ, DMA 4101 EZ, DMA 4501 EZ, DMA 5001 EZ.
DMA 501 EZ
Type-approved three-digit density meter for quick at-line quality checks.
DMA 1001 EZ
Type-approved four-digit density meter for compliant lab results.
DMA 4101 EZ
Type-approved density meter for fast QC in regulated industries.
DMA 4501 EZ
Type-approved reference for production labs.
DMA 5001 EZ
Type-approved benchmark for R&D and standards.
Specifications
| Specification | DMA 501 EZ | DMA 1001 EZ | DMA 4101 EZ | DMA 4501 EZ | DMA 5001 EZ |
|---|---|---|---|---|---|
| Patents | EP3012612B1, AT520632B1, US10145771B2 | EP3012612B1, AT520632B1, US10145771B2 | AT 516420 B1 | AT 516420 B1 | AT 516420 B1 |
| Measuring Range | |||||
| Density | 0.6 g/cm³ to 2 g/cm³ | ||||
| Temperature | 15 °C to 40 °C (59 °F to 104 °F) | 15 °C to 60 °C (59 °F to 140 °F) | 15 °C to 70 °C (59 °F to 158 °F) | 15 °C to 70 °C (59 °F to 158 °F) | 15 °C to 70 °C (59 °F to 158 °F) |
| Pressure | ambient pressure | ||||
| Accuracy* | |||||
| Density | 0.001 g/cm³ | 0.0005 g/cm³ (0.6 g/cm³ to 1.5 g/cm³) 0.001 g/cm³ (1.5 g/cm³ to 2.0 g/cm³) | 0.0005 g/cm³ | 0.0002 g/cm³ | 0.0001 g/cm³ |
| Temperature | 0.3 °C (0.5 °F) | 0.03 °C (0.05 °F)** | 0.02 °C (0.04 °F) | 0.015 °C (0.03 °F) | 0.01 °C (0.02 °F) |
| Repeatability, s.d.*** | |||||
| Density | 0.0002 g/cm³ | 0.00005 g/cm³ | 0.00005 g/cm³ | 0.00005 g/cm³ | 0.000005 g/cm³ |
| Temperature | 0.1 °C (0.2 °F) | 0.02 °C (0.04 °F) | 0.02 °C (0.04 °F) | 0.015 °C (0.03 °F) | 0.01 °C (0.02 °F) |
| Reproducibility, s.d.*** | |||||
| Density | 0.0004 g/cm³ | 0.00007 g/cm³ | 0.00005 g/cm³ | 0.000005 g/cm³ | 0.000005 g/cm³ |
| Digital Resolution | |||||
| Temperature | 0.01 °C (0.01 °F) | 0.01 °C (0.01 °F) | 0.01 °C (0.01 °F) | 0.01 °C (0.01 °F) | 0.001 °C (0.001 °F) |
| Density | 0.0001 g/cm³ | 0.00001 g/cm³ | 0.0001 g/cm³ | 0.00001 g/cm³ | 0.000001 g/cm³ |
| Features | |||||
| Power | U-View, FillingCheck, Full-range viscosity correction | ||||
| Special Functions | Built-in ambient pressure sensor, condition monitoring (humidity) | Built-in ambient pressure sensor, condition monitoring (humidity) | Temperature scan, built-in ambient pressure sensor, condition monitoring | Temperature scan, built-in ambient pressure sensor, condition monitoring | Temperature scan, built-in ambient pressure sensor, condition monitoring |
| Filling Options | Peristaltic pump, Filling kit for pasty samples | Peristaltic pump, Filling kit for pasty samples | Automatic sample changers | Automatic sample changers | |
| Upgrades and Calibrations | Printers, ISO calibration, AP Connect lab execution system | ||||
| Technical Data | |||||
| Ambient Temperature | 5 °C to 35 °C (41 °F to 95 °F) | 5 °C to 35 °C (41 °F to 95 °F) | 15 °C to 35 °C (59 °F to 95 °F) | 15 °C to 35 °C (59 °F to 95 °F) | 15 °C to 35 °C (59 °F to 95 °F) |
| Minimum Sample Volume | Approx. 1 mL | ||||
| Wetted Materials | Borosilicate glass, PTFE, ETFE | ||||
| Dimensions (L x W x H) | 375 mm x 280 mm x 180 mm (14.8 in x 11.0 in x 7.0 in) | 375 mm x 280 mm x 180 mm (14.8 in x 11.0 in x 7.0 in) | 526 mm x 347 mm x 230 mm (20.7 in x 13.7 in x 9 in) | 526 mm x 347 mm x 230 mm (20.7 in x 13.7 in x 9 in) | 526 mm x 347 mm x 230 mm (20.7 in x 13.7 in x 9 in) |
| Weight | 13.5 kg (29.8 lb) | 13.5 kg (29.8 lb) | 22.04 kg (48.6 lbs) | 22.04 kg (48.6 lbs) | 22.04 kg (48.6 lbs) |
| Power Supply | AC 100 to 240 V; 47 to 63 Hz; DC 24V, 3A | AC 100 to 240 V; 47 to 63 Hz; DC 24V, 3A | AC 100 to 240 V, 50/60 Hz, fluctuation ±10 %, 190 VA | AC 100 to 240 V, 50/60 Hz, fluctuation ±10 %, 190 VA | AC 100 to 240 V, 50/60 Hz, fluctuation ±10 %, 190 VA |
| Display | 7 inches, TFT WVGA (800 x 480 Px); PCAP touchscreen | 7 inches, TFT WVGA (800 x 480 Px); PCAP touchscreen | 10.1" TFT WXGA (1280 x 800 px); PCAP touchscreen | 10.1" TFT WXGA (1280 x 800 px); PCAP touchscreen | 10.1" TFT WXGA (1280 x 800 px); PCAP touchscreen |
| Controls | Touchscreen, optional keyboard, mouse, and barcode reader | ||||
| Communication Interfaces | 1 x Ethernet, 3 x USB, 1 x RS232 | 1 x Ethernet, 3 x USB, 1 x RS232 | 4 x USB, Ethernet, CAN, RS232 | 4 x USB, Ethernet, CAN, RS232 | 4 x USB, Ethernet, CAN, RS232 |
| Internal Storage | 25000 measured results including camera images | 25000 measured results including camera images | 60000 measuring values including camera images | 60000 measuring values including camera images | 60000 measuring values including camera images |
| Notes | * According to type approval ** Valid at ± 2 °C ambient temperature recorded during adjustment *** Under ideal conditions and for low densities/viscosities |