Modular Benchtop Density Meter: DMA 4002, DMA 5002, and DMA 6002

The DMA series modular benchtop density meters combine automated functions with unmatched precision and reliability. As the fastest, most modular benchtop density meters in Anton Paar's portfolio, they deliver four-digit results in just 20 seconds.

Anton Paar Modular Benchtop Density Meter: DMA 4002, DMA 5002, and DMA 6002 — Density Meters

Manufacturer: Anton Paar — Austria

Overview

Inspired by over 55 years of expertise in benchtop density measurement, the DMA 4002, DMA 5002, and DMA 6002 modular benchtop density meters combine automated functions with unmatched precision and reliability. As the fastest, most modular benchtop density meters in Anton Paar's portfolio, they deliver four-digit results in just 20 seconds – powered by U-Pulse technology based on the patented Pulsed Excitation Method for best-in-class accuracy.

The instruments offer storage of up to 10,000 measurements and deliver accuracy up to six digits. Daily operations are simplified with features such as U-Dry, one-touch measurement, and a measurement status light, while compliance with a wide range of standards ensures peace of mind.

Outstanding Features that Simplify Workflows

  • One-touch measurement: Measurement at the push of a button
  • U-Dry: Effortless hand-swipe measurement drying
  • FillingCheck: Detection of microbubbles in seconds
  • U-View: Zoomable image of the measuring cell
  • ThermoBalance: Automatic compensation of temperature effects
  • Guided user workflows for simplified operation

Unmatched Precision, Powered by U-Pulse Technology

U-Pulse, based on Anton Paar's patented Pulsed Excitation Method, delivers best-in-class performance – with outstanding precision, repeatability, and reproducibility, whether in high-precision R&D or routine quality control.

Multiparameter Measurement Systems

The DMA 4002, DMA 5002, and DMA 6002 modular benchtop density meters can be connected to a variety of Anton Paar measuring modules, enabling users to capture all key quality control parameters from a single sample – under identical conditions.

Automation that Drives Productivity

Fully automated filling, measuring, and cleaning processes take throughput to the next level. Every Anton Paar benchtop density meter is compatible with a wide range of sample changers, enabling tailored setups for maximum speed and reliability. The instruments ensure accurate, operator-independent results while protecting users by minimizing contact with hazardous substances – all while keeping workflows in motion.

Compliance with Key Standards

The DMA 4002, DMA 5002, and DMA 6002 modular benchtop density meters ensure full compliance with ASTM D4052, D5002, D1250, and ISO 12185 for petroleum testing, as well as all relevant pharmacopeias. They also support data integrity and traceability in line with regulations such as 21 CFR Part 11.

Models

All three models offer one-touch measurement at the push of a button, U-Dry effortless hand-swipe drying, a syringe and measurement status light, combinability with 30+ modules for a measurement system, and automated filling and rinsing via the Xsample series sample changers.

DMA 4002

Results with four-digit accuracy in 20 s.

DMA 5002

Results with five-digit accuracy in 30 s.

DMA 6002

Results with six-digit accuracy for QC and R&D.

Specifications

SpecificationDMA 4002DMA 5002DMA 6002
PatentsAT 516420 B1
Measuring Range
Density0 g/cm³ to 3 g/cm³
Temperature0 °C to 100 °C (32 °F to 212 °F)
PressureUp to 10 bar (145 psi) absolute pressure
Accuracy*
Density0.00005 g/cm³0.00001 g/cm³ (0 g/cm³ to 1.05 g/cm³, 15 °C to 20 °C)
0.00005 g/cm³ (full range)
0.000005 g/cm³
Temperature0.02 °C (0.04 °F)0.01 °C (0.02 °F) (15 °C to 20 °C)
0.015 °C (0.03 °F) (full range)
0.01 °C (0.02 °F)
Repeatability s.d.**
Density0.00001 g/cm³0.000003 g/cm³0.000001 g/cm³
Temperature0.02 °C (0.04 °F)0.005 °C (0.01 °F)0.001 °C (0.002 °F)
Reproducibility s.d.**
Density0.00005 g/cm³0.000005 g/cm³0.000005 g/cm³
Digital Resolution
Density0.00001 g/cm³0.000005 g/cm³0.000001 g/cm³
Features
PowerU-Dry, U-View, U-Pulse, FillingCheck, Status Light, Syringe Illumination, ThermoBalance, Full-range viscosity correction
Special FunctionsTemperature scan, built-in ambient pressure sensor, condition monitoring
Filling OptionsAutomatic sample changers
ModularityAlcohol, pH, color, turbidity, viscosity, refractive index, optical rotation, CO₂, O₂
Upgrades and CalibrationsPrinters, ISO calibration, AP Connect lab execution software
Integrated Tables and FunctionsDensity, specific gravity (SG), alcohol tables, sugar/extract tables, various acid/base tables, API functions
Technical Data
Ambient Temperature15 °C to 35 °C (59 °F to 95 °F)
Typical Measuring Time/Sample20 s30 s40 s
Minimum Sample VolumeApprox. 1 mL
Wetted MaterialsBorosilicate glass, PTFE, ETFE
Dimensions (L x W x H)526 mm x 347 mm x 230 mm (20.7 in x 13.7 in x 9 in)
Weight22.04 kg (48.6 lbs)
Power SupplyAC 100 to 240 V, 50/60 Hz, fluctuation ± 10 %, 190 VA
Display10.1" TFT WXGA (1280 x 800 px); PCAP touchscreen
ControlsTouchscreen, optional keyboard, mouse, and barcode reader
Communication Interfaces4 x USB, Ethernet, CAN, RS232
Internal Storage10,000 measuring values including camera images
Recommended Instrument for Inline MeasurementL-Dens 7400L-Dens 7500L-Dens 7500
Notes* According to ISO 5725
** Under ideal conditions and for low viscosities

Interested in this model? Let's talk specifics.

Request a formal quotation or a detailed technical discussion — our engineers respond within one business day.

Request a Quote

← Density Meters